TN 750

 

The TN750 thermometer offers very high precision measurements to an accuracy of  +/-0.002°C or better (when used with ITS-90 Pt100 –Pt25 optionally- sensor) with resolution at an impressive 0.001°C (over the entire range -180 to +650°C). In addition its self-adaptative linearization mechanism allows to account for probe characteristics automatically during calibration by a National Laboratory (with a residual non-linearity better than 0.001°C). Other key performances are:

§          Long-term drift < 0.001°C +9ppm of Rdg (per year)

§          Short-term stability (jitter) < 0.3 digit ( = 0.0003°C)

§          Temperature coefficient < 1 ppm/°C (max.)

 

 

 

 

 

 

 

Technical specifications

 

 

 

 

 

q  Accuracy

 

Accuracy (after calibration at two fixed points)

0.001°C

 

 

 

 

q  Measurement range

 

Temperature measurement range

-180 to +650 °C (-292 to +1202°F)

Calibrator range

0 to 333 W

 

 

 

 

q  Connection between probe and thermometer

 

Probe type

strain free” Pt100, ITS-90

alpha 1.385 or 1.3926

Optionally: Pt25

Measurement current

1 mA DC

Connection

4-wire shielded cable

Lead wire effect

0.00001°C/W (10W per lead)

NMRR

> 120 dB (at 50 Hz and above)

CMRR

> 140 dB (from 0 to 60 Hz)

 

 

 

 

q  Resolution

 

Thermometer resolution

0.001°C

Calibrator resolution

0.0005W  (660,500 points)

 

 

 

 

q  Probe linearization

 

Linearization characteristics

Based on ITS-90

Self-adaptative & automatic

Residual non-linearity

< 0.001°C

 

 

 

 

q  Stability performances

 

Long-term drift

< 0.001°C +9ppm of Rdg (per year)

Short-term stability (jitter)

0.3 LSD ( = 0.0003°C)

 

 

 

 

q  Auto-zero

 

Auto-zero residual

< 0.0002°C/°C (at the 0°C point)

 

 

 

 

q  Temperature coefficient

 

Temperature coefficient

< 1 ppm/°C (max.)

 

 

 

 

q  Warm-up

 

Warm-up time

15 min. for optimum accuracy

 

 

 

 

q  Adjustment

 

Adjustment of system Zero & Full-Scale

By means of 3 front-panel keys

 

 

 

 

q  Display

 

Display

Easy-to read display with 7 digits

14.2 mm red LEDs

 

 

 

 

q  Communication

 

Communication

RS232 serial mode (ASCII-characters) , isolated command circuit

Optionally: JBUS

 

 

 

 

q  Power

 

Power

110-220VAC,  50-60Hz,  1Watt

 

 

 

 

q  Physical characteristics

 

Case

Extruded aluminium case (EMC)

Full electrical continuity

Version

Laboratory version

(4 feet inc. 2xtip-up feet)

Dimensions

9.1" x 3.3" x 7.5" (inches)

(mm: 230 x 85 x 190 )

Weight

4.4 lb (2 kg)

 

 

 

 

q  Environmental conditions

 

Operating temperature range

0 to 50°C