TN
701R |
|
|
|
The TN701R differential
thermometer has two inputs enabling both direct and differential
temperature measurement. It offers precision measurements to an accuracy of +/-0.01°C or
better (when used with ITS-90 Pt100 or Pt25 sensor) with resolution at an
impressive 0.01°C (over the entire range -180 to +650°C). In addition the
self-adaptative linearization mechanism allows to
account for probe characteristics automatically during calibration by a
National Laboratory (with a residual non-linearity better than 0.002°C).
Other key
performances are: - Long-term
drift < 0.01°C +25ppm of Rdg (per year) - Short-term
stability (jitter) < 0.2 digit ( = 0.002°C) - Temperature
coefficient < 2 ppm/°C (max.) |
|
|
|
|
Technical specifications |
|
|
|
|
|
|
|
|
q Accuracy |
|
|
Accuracy (after calibration at two fixed points) |
0.01°C |
|
|
|
|
|
|
|
q Measurement range |
|
|
Temperature measurement range |
|
|
|
|
|
|
|
|
q Connection between probe and
thermometer |
|
|
Probe type |
2 x Pt100 (1 per channel), ITS90 |
|
Measurement current |
1 mA DC |
|
Connection |
4-wire (with shield) |
|
Lead wire effect |
0.0001°C/ W (10 W per lead) |
|
NMRR |
> 120 dB (at 50 Hz and above) |
|
CMRR |
> 140 dB (from 0 to 60 Hz) |
|
|
|
|
|
|
|
q Resolution |
|
|
Thermometer resolution |
0.01°C |
|
|
|
|
|
|
|
q Input channels |
|
|
Number of measurement channels |
2 x Direct temperature measurements Differential temperature measurement |
|
Channel selection |
With 1 key and 3 [STATUS] red LEDs |
|
|
|
|
|
|
|
q Probe linearization |
|
|
Linearization characteristics |
Based on ITS-90 Self-adaptative & automatic |
|
Residual non-linearity |
< 0.002°C |
|
|
|
|
|
|
|
q Stability performances |
|
|
Long-term drift |
< 0.01°C +25ppm of Rdg (per year) |
|
Short-term stability (jitter) |
0.2 LSD ( = 0.002°C) |
|
|
|
|
|
|
|
q Auto-zero |
|
|
Auto-zero residual |
< 0.0005°C/°C (at the |
|
|
|
|
|
|
|
q Temperature coefficient |
|
|
Temperature coefficient |
<2 ppm/°C (max.) |
|
|
|
|
|
|
|
q Warm-up |
|
|
Warm-up time |
15 min. for optimum accuracy |
|
|
|
|
|
|
|
q Adjustment |
|
|
Adjustment of system Zero & Full-Scale (2 channels) |
By means of 3 front-panel keys |
|
Channel selection |
Using 1 key and 3 [STATUS] red LEDs |
|
|
|
|
|
|
|
q Display |
|
|
Display |
Easy-to read display with 6 digits |
|
|
|
|
|
|
|
q Communication |
|
|
Communication |
RS232 serial
mode (ASCII-characters) , isolated command circuit |
|
|
|
|
|
|
|
q Power |
|
|
Power |
110-220VAC, 50-60Hz, 1Watt |
|
|
|
|
|
|
|
q Physical characteristics |
|
|
Case |
Extruded aluminium case (EMC) Full electrical continuity |
|
Version |
Laboratory version ( |
|
Dimensions |
9.1" x 3.3" x 7.5" (inches) (mm: 230 x 85 x 190 ) |
|
Weight |
|
|
|
|
|
|
|
|
q Environmental conditions |
|
|
Operating temperature range |
0 to |
|
|
|