TN 701R

 

The TN701R differential thermometer has two inputs enabling both direct and differential temperature measurement. It offers precision measurements to an accuracy of  +/-0.01°C or better (when used with ITS-90 Pt100 or Pt25 sensor) with resolution at an impressive 0.01°C (over the entire range -180 to +650°C). In addition the self-adaptative linearization mechanism allows to account for probe characteristics automatically during calibration by a National Laboratory (with a residual non-linearity better than 0.002°C).

Other key performances are:

- Long-term drift < 0.01°C +25ppm of Rdg (per year)

- Short-term stability (jitter) < 0.2 digit ( = 0.002°C)

- Temperature coefficient < 2 ppm/°C (max.)

 

 

 

 

 

 

 

Technical specifications

 

 

 

 

 

q  Accuracy

 

Accuracy (after calibration at two fixed points)

0.01°C

 

 

 

 

q  Measurement range

 

Temperature measurement range

-180 to +650 °C (-292 to +1202°F)

 

 

 

 

q  Connection between probe and thermometer

 

Probe type

2 x Pt100 (1 per channel), ITS90

Measurement current

1 mA DC

Connection

4-wire (with shield)

Lead wire effect

0.0001°C/ W (10 W per lead)

NMRR

> 120 dB (at 50 Hz and above)

CMRR

> 140 dB (from 0 to 60 Hz)

 

 

 

 

q  Resolution

 

Thermometer resolution

0.01°C

 

 

 

 

q  Input channels

 

Number of measurement channels

2 x Direct temperature measurements

Differential temperature measurement

Channel selection

With 1 key and 3 [STATUS] red LEDs

 

 

 

 

q  Probe linearization

 

Linearization characteristics

Based on ITS-90

Self-adaptative & automatic

Residual non-linearity

< 0.002°C

 

 

 

 

q  Stability performances

 

Long-term drift

< 0.01°C +25ppm of Rdg (per year)

Short-term stability (jitter)

0.2 LSD ( = 0.002°C)

 

 

 

 

q  Auto-zero

 

Auto-zero residual

< 0.0005°C/°C (at the 0°C point)

 

 

 

 

q  Temperature coefficient

 

Temperature coefficient

<2 ppm/°C (max.)

 

 

 

 

q  Warm-up

 

Warm-up time

15 min. for optimum accuracy

 

 

 

 

q  Adjustment

 

Adjustment of system Zero & Full-Scale (2 channels)

By means of 3 front-panel keys

Channel selection

Using 1 key and 3 [STATUS] red LEDs

 

 

 

 

q  Display

 

Display

Easy-to read display with 6 digits

14.2 mm red LEDs

 

 

 

 

q  Communication

 

Communication

RS232 serial mode (ASCII-characters) , isolated command circuit

 

 

 

 

q  Power

 

Power

110-220VAC,  50-60Hz,  1Watt

 

 

 

 

q  Physical characteristics

 

Case

Extruded aluminium case (EMC)

Full electrical continuity

Version

Laboratory version

(4 feet inc. 2xtip-up feet)

Dimensions

9.1" x 3.3" x 7.5" (inches)

(mm: 230 x 85 x 190 )

Weight

4.4 lb (2 kg)

 

 

 

 

q  Environmental conditions

 

Operating temperature range

0 to 50°C