TN
600S |
|
|
|
TN600S is a data-logger thermometer
and implements statistical functions
(min., max., average, filtering, recording...). TN600S offers
precision measurements to an accuracy of +/-0.01°C or better (when used with
ITS-90 Pt100 sensor) with resolution at an impressive 0.01°C (over the entire
range -180 to +650°C). In addition its self-adaptative
linearization mechanism allows to account for probe characteristics
automatically during calibration by a National Laboratory (with a residual
non-linearity better than 0.002°C). Other key performances are: - Long-term
drift < 0.01°C +30ppm of Rdg (per year) - Short-term
stability (jitter) < 0.2 digit ( = 0.002°C) - Temperature
coefficient < 2 ppm/°C (max.) |
|
|
|
|
Technical specifications |
|
|
|
|
|
|
|
|
q Accuracy |
|
|
Accuracy (after calibration at two fixed points) |
0.01°C |
|
|
|
|
|
|
|
q Measurement range |
|
|
Temperature measurement range |
|
|
|
|
|
|
|
|
q Connection between probe and
thermometer |
|
|
Probe type |
“strain free” Pt100, ITS-90 alpha 1.385 |
|
Measurement current |
1 mA DC |
|
Connection |
4-wire shielded cable |
|
Lead wire effect |
0.0001°C/W (10W per lead) |
|
NMRR |
> 120 dB (at 50 Hz and above) |
|
CMRR |
> 140 dB (from 0 to 60 Hz) |
|
|
|
|
|
|
|
q Resolution |
|
|
Thermometer resolution |
0.01°C |
|
|
|
|
|
|
|
q Probe linearization |
|
|
Linearization characteristics |
Based on ITS-90 Self-adaptative & automatic |
|
Residual non-linearity |
< 0.005°C |
|
|
|
|
|
|
|
q Stability performances |
|
|
Long-term drift |
< 0.01°C +30ppm of Rdg (per year) |
|
Short-term stability (jitter) |
0.2 LSD ( = |
|
|
|
|
|
|
|
q Auto-zero |
|
|
Auto-zero residual |
< 0.001°C/°C (at the |
|
|
|
|
|
|
|
q Temperature coefficient |
|
|
Temperature coefficient |
< 2 ppm/°C (max.) |
|
|
|
|
|
|
|
q Adjustment |
|
|
Adjustment of system Zero & Full-Scale |
By means of 3 front-panel keys |
|
|
|
|
|
|
|
q Data-logging & statistical
functions |
|
|
STATUS functions |
Min/Max/Average 1 key: [STAT] |
|
FILTER function |
Digital filter 1 key: [FILT] |
|
RECORDING function |
1 to 20 measurements with adjustable time interval (1 to 65535 sec) 1 key : [MEM] |
|
|
|
|
|
|
|
|
|
|
q Display |
|
|
Display |
Easy-to read display with 6 digits |
|
|
|
|
|
|
|
q Communication |
|
|
Communication |
RS232 serial
mode (ASCII) |
|
|
|
|
|
|
|
q Power |
|
|
Power |
110-220VAC, 50-60Hz, 1Watt |
|
|
|
|
|
|
|
q Physical characteristics |
|
|
Case |
Extruded aluminium case (EMC) Full electrical continuity |
|
Version |
Desktop ( |
|
Dimensions |
9.1" x 3.3" x 7.5" (inches) (mm: 230 x 85 x 190 ) |
|
Weight |
|
|
|
|
|
|
|
|
q Environmental conditions |
|
|
Operating temperature range |
0 to |
|
|
|