Range |
-180.00...
+655.00 °C (-292 to 1221.8 °F) |
|
| |
Resolution |
0.01°C |
|
| |
Sensor
type |
Pt100
(4-wire with shield)
European (DIN) or US
Industrial Standard |
|
| |
Measurement
current |
1
mA DC |
|
| |
Lead
wire effect |
0.0001°C/ohm,
10 ohm per lead |
|
| |
Linearity |
0.005°C |
|
| |
Long-term
drift |
0.01°C
+50 ppm of reading per year |
|
| |
Short-term
stability (jitter) |
0.2
digit ( = 0.002°C) |
|
| |
Temperature
coefficient |
4
ppm/°C (max.) |
|
| |
Power |
PB
- 6V 2.8Ah Battery (rechargeable, battery charger
supplied) |
|
| |
Autonomy |
60
hours |
|
| |
Dimensions |
inches:
9.1" x 3.3" x 7.5"
mm:
230 x 85 x 190 |
|
| |
Case |
Extruded
aluminium case (CE marked) |
|
| |
Weight |
4.4
lb (2 kg) |
|
| |
Version |
Portable
and Laboratory, 4 feet (inc. 2xtip-up feet) |
|
| |
Display |
6
digits, 14.2 mm red LEDs |
|
| |
Adjustments |
Zero-point
and Gain (3 keys) |
|
| |
STATUS
function |
Min/Max/Average
(1 key: [STAT]) |
|
| |
FILTER
function |
Digital
filter (1 key: [FILT]) |
|
| |
RECORDING
function |
1
to 20 measurements with adjustable time interval
(1 to 65535 sec., 1 key : [MEM]) |
|
| |
READ
function |
Reading
of: interval (time), 20 last recorded
measurements, Min, Max, Average, ... |
|
| |
AUTO-OFF
function |
Auto-off
after 5 minutes |
|
| |
BATTERY-TEST
function |
At
5.4 V: flashes then stops after 120 sec. |
|